Surtron Flash 200

SURTRON 400 HP can be used to simultaneously perform monopolar and bipolar surgery in application fields which require high precision and reliability.

DOUBLE ACTIVITY ‘OPERATION
• Two output channels for monopolar cutting
• Two independent setting and output channels for monopolar coagulation more frequently

CONTROLLED CUTTING PROCEDURES
• cutting currents for eschar free incision
• Cutting current with selectable coagulating property for bleeding controlled dissection
• Low capacitive coupling by current endoscopic procedures Slow BLEND

DOUBLE PROCEDURES SUBSIDIARIES COAGULATION
• SPEEDY Coag Coag DEEP ** * and two operators
• SPRAY COAG achievable with handpiece or pedal

WIDE BIPOLAR FEATURES
• Bipolar cutting with bleeding control
• Cutting current Slow BLEND
• Start and Stop automatic activated separately

TOTAL SAFETY FOR PATIENT AND SURGEON
• Acceptability of contact impedance between the patient’s body and the neutral plate constantly verified by the electronic control Skin Plate Electronic Control called SPEC
• Continuous monitoring of leakage current
• Change in automatically compensated mains voltage before starting the RF power output

CUSTOM PROCEDURES
• Independent choice of coagulation currents and power levels of each of the two operators
• Ten configurable personal procedures and immediately recallable

CONDUCT OF EXTENSIONS AND SIGNAL CONNECTION
• outputs for connections driven argon plasma units and smoke aspirator

AUTOMATIC SELF-DIAGNOSIS WITH SURTRON 400 HP
• Automatic diagnostic control of the actual complete electronic drive functionality and accessories connected and coded report of found failures

* Coag SPEEDY is also called Fulgurate or Forced
** Coag DEEP is also called Pin Point, Dessicate or Soft

FEATURES
Code 10400.902
Max output power monopolar pure CUT
400 W – 300 Ω
Max output power monopolar cut-coag CUT80%
300 W – 300 Ω
Max output power monopolar cut-coag CUT60%
250 W – 300 Ω
Max output power monopolar cut-coag CUT40%
200 W – 300 Ω
Max output power monopolar cut ENHANCED
250 W – 500 Ω
Max output power monopolar cut-coag BLEND
250 W – 300 Ω
Max output power monopolar coag SPEEDY
120 W – 500 Ω
Max output power monopolar coag DEEP
120 W – 200 Ω
Max output power monopolar coag SPRAY
120 W – 2000 Ω

Max output power monopolar coag SPRAY ARGON
120 W – 2000 Ω
Max output power bipolar CUT
120 W – 50 Ω
Max output power bipolar cut-coag CUT80%
120 W – 50 Ω
Max output power bipolar cut-coag CUT60%
120 W – 50 Ω
Max output power bipolar cut-coag CUT40%
100 W – 50 Ω
Max output power bipolar cut-coag BLEND
120 W – 150 Ω
Max output power bipolar coag COAG
120 W – 50 Ω
Vessel sealing
200 W – 50 Ω

Working frequency Monopolar / Bipolar
425 kHz / 525 kHz
Patient circuit
F
Selectable input voltage
115 – 230 Vac
Mains frequency
50 – 60 Hz
Electrical input power
1000 VA

Size WxHxD
470x150x400 mm
Weight
17,5 Kgs

CONTROLS
Patient/Plate circuit monitoring
Output power monitoring
Self check control
H.F. Leakage currents control
Compensation of mains fluctuation

SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Level of protection provided by housing: IP21
Electrical classification: CF
MDD 93/42 / EC: II b

Categories: , Tag:

SURTRON FLASH 200  is an electrosurgical equipment which is suitable to precision monopolar and bipolar surgery and micro-surgery without tissue alterations.

SURTRON FLASH 200 can supply RF operative powers either for programmable time as short as hundreds of a second.

Through its performances, SURTRON FLASH 200, allows pure CUT,  cut-coagulation BLEND, incision with reduced prodution of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION.

The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job.

SURTRON FLASH 200 allows a highly professional surgery thanks to the user-friendly and safety solutions normally used.

The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

FEATURES
Code 10300.401
Max output power CUT 150 W-500 Ω
Max output power BLEND 100 W-500 Ω
Max output power ENHANCED 100 W-500 Ω
Max output power FORCED COAG 100 W-500 Ω
Max output power SOFT COAG 80 W-100 Ω
Max output power BIPOLAR COAG 60 W-50 Ω
Working frequency 60 W-100 Ω
Working frequency 600 KHz
RF power pulse emission time 4.0 MHz
RF power pulse emission time 10 ms – 30 sec
RF power pulse interval 1 – 9
Patient circuit F
Selectable input voltage 115 – 230 Vac
Mains frequency 50 – 60 Hz
Electrical input power 350 VA
Size WxHxD 370x144x319 mm
Weight 6 Kgs

CONTROLS
Control Plate Circuit / OC patient
Control of the emitted power
Self-diagnosis

SAFETY
EN60601-1
EN60601-1-2
EN60601-2-2
Electrical classification: CF
MDD 93/42 / EC: II b